INVACARE® FREEWAY® FREEDOMINVACARE® FREEWAY® FREEDOM
INVACARE® FREEWAY® FREEDOMINVACARE® FREEWAY® FREEDOM
INVACARE® FREEWAY® FREEDOM
AEROSOL COMPRESSOR
MODEL IRC1175
OPERATOR'S MANUAL
DEALER: THIS MANUAL MUST BE
GIVEN TO THE END USER.
USER: BEFORE USING THIS
COMPRESSOR, READ THIS
MANUAL AND SAVE FOR
FUTURE REFERENCE.
Vue de la page 0
1 2 3 4 5 6 ... 23 24

Résumé du contenu

Page 1 - OPERATOR'S MANUAL

1INVACARE® FREEWAY® FREEDOMINVACARE® FREEWAY® FREEDOMINVACARE® FREEWAY® FREEDOMINVACARE® FREEWAY® FREEDOMINVACARE® FREEWAY® FREEDOMAEROSOL COMPRESSORM

Page 2 - SPECIAL NOTES

10THE FREEWAY FREEDOM BATTERY PACKINITIAL BATTERY CHARGINGTo maximize the capacity and life of the Freeway Freedom rechargeablebatteries, FULLY charge

Page 3 - TABLE OF CONTENTS

11OPERATIONPOWER SUPPLY OPTIONSFreeway Freedom may be operated from three (3) different sources:1. The wall outlet, using the Freeway Freedom AC/DC po

Page 4 - SAFETY SUMMARY

12OPERATIONOPERATION FROM THE WALL OUTLETThe Freeway Freedom power pack is required for running the FreewayFreedom from AC wall outlets and to recharg

Page 5

13OPERATIONCAUTIONIf the unit does NOT operate, IMMEDIATELY disconnectthe power supply from the DC inlet and check for correctpolarity (Refer to SPECI

Page 6 - FEATURES

14OPERATION3. Connect the long hose end of the coiled duratube to the inlet onthe bottom of the Sidestream nebulizer.4. Connect either the mouthpiece

Page 7 - MODEL IRC1175 SPECIFICATIONS

15OPERATIONTAKING THE TREATMENTCAUTIONDO NOT use the Freeway Freedom for more than 20minutes continuously in any one (1) hour. Allow a minimumof 40 mi

Page 8 - TRAVELING INFORMATION

16AFTER THE TREATMENT (FIGURE 3)1. Disconnect the Sidestream nebulizer from the coiled duratube bygently twisting the tubing.2. Rinse out the Sidestre

Page 9 - BATTERY PACK

17OPERATION/MAINTENANCEFIGURE 3 - AFTER THE TREATMENTMAINTENANCEPROLONGING THE LIFE OF YOUR BATTERYThe Freeway Freedoms battery is fully sealed and r

Page 10 - PROCEDURES

18MAINTENANCENOTE: If you are not using the Freeway Freedom for periods longer than one (1) month,remove the battery pack from the device and store se

Page 11 - OPERATION

19MAINTENANCEYour Freeway Freedom, battery pack and power pack can be cleanedby wiping with a damp cloth.The carrying case can be machine washed (40oC

Page 12 - FREEDOM BATTERY PACK

2SPECIAL NOTESDANGER/WARNING/CAUTION notices as used in thismanual apply to hazards or unsafe practices which couldresult in personal injury or proper

Page 13

20FUSE REPLACEMENTThe 12 volt DC lead (cigarette lighter lead) is fitted with a T3 amp fuse.This is located in the cigarette lighter plug and is remov

Page 14

21TROUBLESHOOTINGPROBLEM POSSIBLE SOLUTIONS Unit does not run from battery. 1. Assure battery is installedproperly.2. Remove Protective film (tape)

Page 15 - TAKING THE TREATMENT

22LIMITED WARRANTYPLEASE NOTE: THE WARRANTY BELOW HAS BEENDRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLETO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.Thi

Page 16

23LIMITATIONS AND EXCLUSIONS: THE FOREGOINGWARRANTY SHALL NOT APPLY TO SERIAL NUMBEREDPRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED ORDEFACED, PRODUC

Page 17 - MAINTENANCE

Invacare Corporation www.invacare.comUSAOne Invacare WayElyria, Ohio USA44036-2125800-333-6900699 Aero LaneSanford, Florida32771Technical Support:Tol

Page 18 - CLEANING THE FREEWAY FREEDOM

3TABLE OF CONTENTSTABLE OF CONTENTSSPECIAL NOTES ... 2TABLE OF CONTENTS ...

Page 19

4SAFETY SUMMARYSAFETY SUMMARYCAUTION"Caution: Federal law restricts this device to sale or rentalby or on order of a physician, or any other prac

Page 20 - ACCESSORIES AND SPARE PARTS

5WARNINGDO NOT operate unit longer than twenty (20) minutesCONTINUOUSLY. Allow forty (40) minutes for the unit tocool before operating again.Use this

Page 21 - TROUBLESHOOTING

6NOTE: The 12 volt DC lead, AC/DC power pack and power cord may be stored inthe detachable accessory bag.Child MaskAdult MaskSidestreamNebulizerCoiled

Page 22 - LIMITED WARRANTY

7SPECIFICATIONSElectrical Requirements: 90-240 VAC, 50-60 Hz, or12 -14 VDC Ambient TemperatureRange: 10o C - 40o C (50o F - 104o F)NOTE: Operating at

Page 23

8INTRODUCTION/TRAVELING INFORMATIONINTRODUCTIONFreeway Freedom is different than other aerosol compressors becauseit is smaller, lighter and quieter.

Page 24

9THE FREEWAY FREEDOM BATTERY PACKTHE FREEWAY FREEDOMBATTERY PACKThe Freeway Freedom battery pack contains nickel metal hydride(NiMH) cells. Refer to P